Hioki-CT6843 AC/DC Current probe, frequency bandwidth DC to 500 kHz, 200 A AC/DC
Consistent, High-precision Current Testing Across a Wide Temperature Range
Key Features and Functions
- Ideal for use in environmental testing with broad -40?C to 85?C temperature range
- High precision with a clamp-type design, ±0.3% amplitude accuracy, ±0.1° phase accuracy
- Wide-bandwidth DC to 500 kHz
- Single-handed operation and robust locking mechanism
- Reduced effects from magnetic fields, conductor position, and noise from nearby wires
- For EV/HEV battery charge and discharge efficiency measurement and inverter and power conditioner conversion efficiency evaluation
Documents and Software:
Availability:
Call or email for delivery time
Hioki AC/DC clamp sensors are best-in-class current sensors for use with power meters, Memory HiCorders and high performance oscilloscopes. The CT6843 is DC to 500 kHz/200 A sensors, ideal for capturing signals during the efficiency testing of batteries, inverters and power conditioners.
Key Specifications:
- Rated current: 200 A AC/DC
- Max. allowable input: 400 A rms (requires derating)
- Frequency characteristics: Amplitude: DC to 500 kHz
- Phase: DC to 300 kHz
- Amplitude and phase accuracy:
- DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined)
- DC< f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1°) Defined to 500 kHz
- Power consumption: 6 VA max. (at 200 A/55 Hz, ±12 V power requirement)
- Output voltage:
- 0.1 V/A (CT6841), 0.01 V/A (CT6843)
(voltage output with the Sensor Unit, use with a device having a 1 MΩ input resistance or higher) - Core diameter: φ 20 mm (0.79 in)
Description
Additional Informations
Reviews
Query on the Item
Do you have any questions on this item?
Description
Reviews
Query on the Item
Hioki AC/DC clamp sensors are best-in-class current sensors for use with power meters, Memory HiCorders and high performance oscilloscopes. The CT6843 is DC to 500 kHz/200 A sensors, ideal for capturing signals during the efficiency testing of batteries, inverters and power conditioners.
Key Specifications:
- Rated current: 200 A AC/DC
- Max. allowable input: 400 A rms (requires derating)
- Frequency characteristics: Amplitude: DC to 500 kHz
- Phase: DC to 300 kHz
- Amplitude and phase accuracy:
- DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined)
- DC< f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1°) Defined to 500 kHz
- Power consumption: 6 VA max. (at 200 A/55 Hz, ±12 V power requirement)
- Output voltage:
- 0.1 V/A (CT6841), 0.01 V/A (CT6843)
(voltage output with the Sensor Unit, use with a device having a 1 MΩ input resistance or higher) - Core diameter: φ 20 mm (0.79 in)