Hioki-3153 AUTOMATIC INSULATION/WITHSTANDING HiTESTER
All-in-one Model that Combines Withstand Voltage and Insulation Resistance (AC/DC)
Key Features and Functions
- Programmable insulation (50 to 1,200 V DC) and dielectric strength (AC/DC) testing
- Program up to 32 files of test types, test points (50 steps), and measurement settings
- Optional scanner for multipoint automatic testing
- Uses the PWM method to generate accurate test voltages that do not depend on the supply voltage
- Ramp timer function for increasing or decreasing the applied voltage during dielectric strength testing at user-specified times
Documents and Software:
Availability:
Call or email for delivery time
The Hioki electrical safety testers are designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. The 3153 is an automatic insulation and AC/DC voltage withstanding tester that fully supports automated testing.
Key Specifications:
Withstanding test
- Testing voltage AC/DC: 0.2 kV to 5.00 kV AC, 500 VA (max. 30 minutes), 0.2 kV to 5.00 kV DC, 50 VA (continuance)
- Voltage setting: Digital setting (0.01 kV setting resolution)
- Waveform/Frequency: Sine wave (5% or less distortion, unloaded), 50/60 Hz selectable
- Current measurement: 0.01 mA to 100.0 mA, Average rectified display (Digital)
- Measurement range: 10 mA (0.01 mA resolution), 100 mA (0.1 mA resolution)
- Voltmeter: Digital: accuracy ±1.5 % f.s. (f.s.=5.00 kV) (Average rectified display)
- Decision method: Window comparison (digital settings)
Insulation test
- Rated testing voltage: 50 to 1,200 V DC (in 1 V steps)
- Rated testing current: 1 mA, Short-circuit current: 200 mA or less
- Measurement range/accuracy: 0.10 to 9999 MΩ, 4 ranges, ± 4 % rdg. (representative values for 0.5 MΩ to 1,000 MΩ)
- Decision method: Window comparison (digital settings)
Description
Additional Informations
Reviews
Query on the Item
Do you have any questions on this item?
Description
Reviews
Query on the Item
The Hioki electrical safety testers are designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. The 3153 is an automatic insulation and AC/DC voltage withstanding tester that fully supports automated testing.
Key Specifications:
Withstanding test
- Testing voltage AC/DC: 0.2 kV to 5.00 kV AC, 500 VA (max. 30 minutes), 0.2 kV to 5.00 kV DC, 50 VA (continuance)
- Voltage setting: Digital setting (0.01 kV setting resolution)
- Waveform/Frequency: Sine wave (5% or less distortion, unloaded), 50/60 Hz selectable
- Current measurement: 0.01 mA to 100.0 mA, Average rectified display (Digital)
- Measurement range: 10 mA (0.01 mA resolution), 100 mA (0.1 mA resolution)
- Voltmeter: Digital: accuracy ±1.5 % f.s. (f.s.=5.00 kV) (Average rectified display)
- Decision method: Window comparison (digital settings)
Insulation test
- Rated testing voltage: 50 to 1,200 V DC (in 1 V steps)
- Rated testing current: 1 mA, Short-circuit current: 200 mA or less
- Measurement range/accuracy: 0.10 to 9999 MΩ, 4 ranges, ± 4 % rdg. (representative values for 0.5 MΩ to 1,000 MΩ)
- Decision method: Window comparison (digital settings)